Line 9: Line 9:
    * Small dilution refrigerators ("la Jolla" style) have diameters of a few cm, their 3He flow rate is ~10 to 50 µmol/sec. They are often used as "inserts" in a liquid helium dewar . Due to their small size, the heat exchangers have a small total area, and heat leaks make it difficult to reach temperatures below 10 mK.    * Small dilution refrigerators ("la Jolla" style) have diameters of a few cm, their 3He flow rate is ~10 to 50 µmol/sec. They are often used as "inserts" in a liquid helium dewar . Due to their small size, the heat exchangers have a small total area, and heat leaks make it difficult to reach temperatures below 10 mK.
  
-   * Larger refrigerators with heat exchangers of typical diameters larger than 10 cm ("Grenoble design") can reach temperatures on the order of 2 mK, with flow rates of 100 *mol/s. They can also provide large cooling power, with flow rates up to 10 mmol/s, obviously at higher working temperatures. +   * Larger refrigerators with heat exchangers of typical diameters larger than 10 cm ("Grenoble design") can reach temperatures on the order of 2 mK, with flow rates of 100 µmol/s. They can also provide large cooling power, with flow rates up to 10 mmol/s, obviously at higher working temperatures. 
    * Cooling power at the MC is given by the simple expression    * Cooling power at the MC is given by the simple expression
  
Line 49: Line 49:
  
   * The still plays an important role: the **3He flow rate** of the dilution refrigerator, and hence the cooling power available at the level of the mixing chamber, **is mainly determined by the power applied to the still**.      * The still plays an important role: the **3He flow rate** of the dilution refrigerator, and hence the cooling power available at the level of the mixing chamber, **is mainly determined by the power applied to the still**.   
 +
 +===== The Continuous heat exchanger  =====
 +
 +===== The Step ("discrete") heat exchangers =====
  
 ===== The Mixing Chamber  ===== ===== The Mixing Chamber  =====
Line 73: Line 77:
  
  
-===== Troubleshooting =====+===== Troubleshooting Dilution Refrigerators =====
   * 1 K pot hot, pressure is low. Filling capillary blocked. Remove LHe so that the bath level is below the intake, keep 4He pressure in Pot above bath pressure. Having a heater on the 1K pot capillary can help, applying heat pulses...   * 1 K pot hot, pressure is low. Filling capillary blocked. Remove LHe so that the bath level is below the intake, keep 4He pressure in Pot above bath pressure. Having a heater on the 1K pot capillary can help, applying heat pulses...
  
wiki/design_construction.1643823092.txt.gz · Last modified: 2022/02/02 17:31 by henri.godfrin@neel.cnrs.fr